The UV200 probe offers a simple way to measure organic matter in surface water or industrial/municipal wastewater. It must be calibrated for each application to give an equivalent COD, BOD or TOC measurement depending on the composition of the sample. It is based on a new light source technology offering a compact and economical solution with a long service life.
| Dimension Drawing | Measuring Range | 0 - 600 Abs/m (equivalent to 0 - 2000 mg/L COD on municipal waste water) 0- 200 Abs/m (equivalent to 0 - 100 mg/L COD on surface water) |
|---|---|---|
![]() |
Accuracy | +/- 2% on standard solution |
| Optical path | 3 or 10 mm | |
| Response time | 10 seconds | |
| Turbidity compensation | Integrated by dual-beam method | |
| Light source life time | > 3 years | |
| Measurement technique | UV light absorbance | |
| Mounting | Immersion | |
| Probe cleaning system | Air cleaning option available | |
| Temperature range | -20˚C to 70˚C | |
| Body material | Stainless Steel 316L | |
| Wet material | SS 316L, PTFE, Quartz, FKM, PE, Neoprene | |
| Connection | Screw terminal | |
| Cable length | 10 meters | |
| Sensor Protection | IP68 |